M400E Ozone Analyzer Operator’s Manual 
Optional Hardware and Software 
5.7.3. ETHERNET (OPT 63) 
The ETHERNET option allows the analyzer to be connected to any Ethernet local area network (LAN) running 
TCP/IP.  The local area network must have routers capable of operating at 10BaseT.  If Internet access is 
available through the LAN, this option also allows communication with the instrument over the public Internet.  
Maximum communication speed is limited by the RS-232 port to 115.2 kBaud. 
When installed, this option is electronically connected to the instrument’s COM2 serial port making that port no 
longer available for RS-232/RS-485 communications.   
The option consists of a Teledyne Instruments designed Ethernet card (see figures below), and a 7-foot long 
CAT-5 network cable, terminated at both ends with standard RJ-45 connectors. 
 
 
Figure 5-8: 
M400E Ethernet Card  
Rear Panel 
(as seen from inside)
 
CPU 
Card
 
Ethernet 
Card 
Female RJ-45 
Connector
 
RS-232 
Connector To 
Motherboard
 
LNK LED
 
ACT LED
 
TxD LED
 
RxD LED
 
 
Interior View 
Exterior View 
 
Figure 5-9: 
M400E Rear Panel with Ethernet Installed 
  For more information on setting up and using this option, see Section 8.4 
04315 Rev. C1 