Agilent Technologies E1441A Benutzerhandbuch

Seite von 206
Agilent E1441A Specifications  
135
Appendix A
Appendix A
Agilent E1441A Specifications
Waveforms
Standard Waveforms:
Sine, Square, Triangle, Ramp, Noise, DC
 
volts, Sine (x)/x, Negative Ramp,
Exponential Rise, Exponential fall, Cardiac
Arbitrary Waveforms:
Waveform Length:
8 to 16,000 points
Amplitude Resolution:
12 bits (including sign)
Sample Rate:
40 MSa / sec
Non-Volatile Memory:
Four (4) 16k waveforms
Frequency
Characteristics
Sine:
100 
µHz - 15 MHz
Square:
100 
µHz - 15 MHz
Triangle:
100 
µHz - 100 kHz
Ramp:
100 
µHz - 100 kHz
Noise (Gaussian):
10 MHz bandwidth
Arbitrary Waveforms:
8 to 8,192 points
100 
µHz - 5 MHz
8,193 to 12,287 points
100 
µHz - 2.5 MHz
12,288 to 16,000 points
100 
µHz to 200 kHz
Resolution:
10 
µHz or 10 digits
Accuracy:
10 ppm in 90 days
20 ppm in 1 year
18
° C - 28°
C
Temperature Coefficient:
<2 ppm/ 
° C
Aging:
<10 ppm/yr
Sinewave Spectral Purity
Harmonic Distortion
DC to 20 kHz:
< -70 dBc
20 kHz to 100 kHz:
< -60 dBc
100 kHz to 1 Mhz:
< -45 dBc
1 Mhz to 15 Mhz:
< -35 dBc
Total Harmonic Distortion
DC to 20 kHz:
< 0.04%
Spurious (non-harmonic)
Output (DC to 1 Mhz):
< -65 dBc
Output ( 1 Mhz):
< -65 dBc + 6 dB/octave
Phase Noise:
< -52 dBc in a 30 kHz band