4-10
Reference
Cleaning Connections for Accurate Measurements
Cleaning Connections for Accurate 
Measurements 
Today, advances in measurement capabilities make connectors and connec-
tion techniques more important than ever. Damage to the connectors on cali-
bration and verification devices, test ports, cables, and other devices can 
degrade measurement accuracy and damage instruments. Replacing a dam-
aged connector can cost thousands of dollars, not to mention lost time! This 
expense can be avoided by observing the simple precautions presented in this 
book. This book also contains a brief list of tips for caring for electrical connec-
tors.
Choosing the Right Connector
A critical but often overlooked factor in making a good lightwave measure-
ment is the selection of the fiber-optic connector. The differences in connec-
tor types are mainly in the mechanical assembly that holds the ferrule in 
position against another identical ferrule. Connectors also vary in the polish, 
curve, and concentricity of the core within the cladding. Mating one style of 
cable to another requires an adapter. Agilent Technologies offers adapters for 
most instruments to allow testing with many different cables. Figure 4-2 on 
page 4-11 s
hows the basic components of a typical connectors.
The system tolerance for reflection and insertion loss must be known when 
selecting a connector from the wide variety of currently available connectors. 
Some items to consider when selecting a connector are: 
• How much insertion loss can be allowed? 
• Will the connector need to make multiple connections? Some connectors are 
better than others, and some are very poor for making repeated connections. 
• What is the reflection tolerance? Can the system take reflection degradation? 
• Is an instrument-grade connector with a precision core alignment required? 
• Is repeatability tolerance for reflection and loss important? Do your specifica-