Agilent Technologies 83491 Benutzerhandbuch

Seite von 60
4-13
Reference
Cleaning Connections for Accurate Measurements
The soft core, while allowing precise centering, is also the chief liability of the 
connector. The soft material is easily damaged. Care must be taken to mini-
mize excessive scratching and wear. While minor wear is not a problem if the 
glass face is not affected, scratches or grit can cause the glass fiber to move 
out of alignment. Also, if unkeyed connectors are used, the nickel silver can be 
pushed onto the glass surface. Scratches, fiber movement, or glass contamina-
tion will cause loss of signal and increased reflections, resulting in poor return 
loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immedi-
ately obvious to the naked eye, poor measurements result without the user 
being aware. Microscopic examination and return loss measurements are the 
best way to ensure good measurements. Good cleaning practices can help 
ensure that optimum connector performance is maintained. With glass-to-
glass interfaces, any degradation of a ferrule or the end of the fiber, any stray 
particles, or finger oil can have a significant effect on connector performance. 
Where many repeat connections are required, use of a connector saver or 
patch cable is recommended.
Figure 4-5 shows the end of a clean fiber-optic cable. The dark circle in the 
center of the micrograph is the fiber’s 125 
µm core and cladding which carries 
the light. The surrounding area is the soft nickel-silver ferrule. Figure 4-6 
shows a dirty fiber end from neglect or perhaps improper cleaning. Material is 
smeared and ground into the end of the fiber causing light scattering and poor 
reflection. Not only is the precision polish lost, but this action can grind off the 
glass face and destroy the connector.
Figure 4-7 shows physical damage to the glass fiber end caused by either 
repeated connections made without removing loose particles or using 
improper cleaning tools. When severe, the damage of one connector end can 
be transferred to another good connector endface that comes in contact with 
the damaged one. Periodic checks of fiber ends, and replacing connecting 
cables after many connections is a wise practice.
The cure for these problems is disciplined connector care as described in the 
following list and in “Cleaning Connectors” on page 4-17.