Agilent 83491/2/3A modules require that firmware revision A.06.25 or later be 
installed in the Agilent 83480A. If you wish to install the module in an 
Agilent 54750A digitizing oscilloscope, you must first install the 
Agilent 83480K communications firmware upgrade kit.
To check the Agilent 83480A’s firmware revision code
1
Press the 
Utility key and then the System config softkey.
2
The firmware revision number is listed under the Frame section of the display.
C A U T I O N
Fiber-optic connectors are easily damaged when connected to dirty or 
damaged cables and accessories. The Agilent 83492A and 83493A front-panel 
input connectors are no exception. When you use improper cleaning and 
handling techniques, you risk expensive instrument repairs, damaged cables, 
and compromised measurements. Before you connect any fiber-optic cable to 
an Agilent 83492A or 83493A module, refer to “Cleaning Connections for 
Accurate Measurements” on page 4-10.
C A U T I O N
The circuits on electrical inputs and outputs can be damaged by electrostatic 
discharge (ESD). Therefore, avoid applying static discharges to any front or 
rear-panel electrical connector. Before connecting any coaxial cable to a front-
panel connector, momentarily short the center and outer conductors of the 
cable together. Avoid touching the front-panel connectors without first 
touching the frame of the instrument. Be sure that the instrument is properly 
earth-grounded to prevent buildup of static charge. Refer to “Electrostatic 
Discharge Information” on page 4-8.