                                  Guidance and manufacturer’s declaration - electromagnetic emissions information  
The Trident Image Intensifier is intended for use in the electromagnetic environment specified below. The 
customer or the user of the Trident Image Intensifier should assure that it is used in such an  
environment.  
IMMUNITY test  IEC 60601 
Compliance level 
Electromagnetic environment - guidance 
Test level 
Electrostatic 
± 6 kV contact 
± 6 kV contact 
Floors should be wood, concrete or ceramic 
Discharge (ESD) 
± 8 kV air 
± 8 kV air 
tile. If floors are contain synthetic 
IEC 61000-4-2 
material, the relative humidity should be at 
least 30%. 
Electrical fast 
± 2kV for power 
Not applicable 
Mains power quality should be that of a 
Transient/burst 
Supply lines 
typical commercial or hospital environment. 
± 1kV for 
IEC 61000-4-4 
input/output lines 
Surge 
± 1 kV line(s) to 
Not applicable 
Mains power quality should be that of a 
IEC 61000-4-5 
line(s) 
typical commercial or hospital environment. 
± 2kV line(s) to 
earth 
Voltage dips, 
<5% U 
Mains power quality should be that of a 
short 
(>95% dip in U 
T
typical commercial or hospital environment. 
interruptions and  for 0.5 cycle) 
If the user of the (ME EQUIPMENT or ME 
voltage variations 
SYSTEM) requires continued operation 
on power supply  40% U 
during power mains interruptions, it is 
input lines 
(60% dip in U 
T
Not applicable 
recommended that the (ME EQUIPMENT or 
for 5 cycles 
ME SYSTEM) be powered from an 
IEC 61000-4-11 
uninterruptible power supply or a battery. 
70% U 
T  
(30% dip in U 
T
)  
for 25 cycles  
 
<5% U 
T  
(95% dip in U 
T
)  
for 5 s  
Power frequency  3 A/m 
Not applicable 
Power frequency magnetic fields should be at 
(50/60 Hz) 
levels characteristic of a typical location in a 
magnetic field 
typical commercial or hospital environment. 
IEC 61000-4-8  
NOTE U 
is the a.c. mains voltage prior to application of the test level.  
 
 
 
 
 
 
 
 
 
 
 
 
 
4