Agilent Technologies 5060-3386 Leaflet

Page of 2
Set the Sense Switch
Each connector has a local/remote sense switch. Unless
you are using remote sensing, make sure that the sense
switch is set to Local. Remote sensing is used in
applications that require voltage sensing directly at the
load.
Installing a Relay Board in a Standard
Connector
You can upgrade a standard connector to a relay
connector by clipping six jumpers in the connector and
installing a relay board (p/n 5060-3351).
Open the Connector
Open the connector by loosening the four corner screws
(refer to the figure on the first page).
Clip the Jumpers on the Connector Board
Use a pair of diagonal cutters and clip the eight jumpers
on the connector board. Make sure the jumpers are cut
before you install the relay board. It is difficult to remove
the relay board once it is installed.
2
1
3
1 RELAY BOARD
2 CUT W603 W604
3 CUT W601 W602 W605 W606 W607 W608
Install the Relay Board
Plug the relay board into the J602 and J603 connectors on
the connector board.
Close the Connector
Close the connector and tighten the four corner screws.
Observe the                  on the figure.
With a marker, check the Relay box on the connector
label to indicate tha a relay board is installed.
1. Relay Board
2. Cut W603 W604
3. Cut W601 W602 W605 W606 W607 W608
CAUTION
   CAUTION
 Hand tighten only.
 Start all screws
   before tightening.
 Tighten screws evenly.
 DO NOT overtighten screws.
 DO NOT use locking-type
   adhesive on the screws.