Philips 862478 User Manual

Page of 182
Preparing to Use the Monitor
2-4 
Site Preparations  
Preparing to Use the Monitor
Warning
Warning
To avoid contaminating or infecting personnel, the service environment or other 
equipment, make sure that equipment which has been used before has 
been appropriately disinfected and decontaminated.
Power 
Source 
Requirements
Protecting 
Against 
Electrical 
Shock
The C3 patient monitor is classified as Class I Equipment with an internal power source 
according to IEC 60601-1/CSAC22.2 601.1/UL 2601-1, which means that it is a monitor 
included in the protective grounding (protective earth) system of the room by way of 
grounding contacts in the power plug.
Warning
Warning
Disconnect the monitor from the AC source by unplugging the power cable from the AC 
power connector located on the rear of the monitor. The On/Standby 
button does not disconnect the monitor from the AC mains supply.
Equipotential
 
Grounding
To eliminate potential differences between difference pieces of equipment, the monitor must 
have a separate connection to the equipotential grounding system.
One end of the equipotential grounding cable (potential equalization conductor) is connected 
to the equipotential grounding post on the rear of the monitor, and the other end is connected 
to one point of the equipotential grounding system.
Combining 
Equipment
All combinations of medical equipment with non-medical equipment must comply with IEC 
60601-1-1.
Warning
Warning
Accessory equipment connected to the monitor’s data interface must be certified 
according to IEC Standard 60950 for data-processing equipment or 
IEC Standard 60601-1 for electromedical equipment. All combinations 
of equipment must be in compliance with IEC Standard 60601-1-1 
systems requirements.