Calculating Cerebral Perfusion
12 Monitoring Invasive Pressure
157
5
Select the calibration pressure from the list, for example 200 mmHg.
6
Select 
Confirm
 to recalculate the calibration factor using the applied pressure. 
7
When the monitor displays <
Press> calibration done at <date and time>
, remove 
the manometer tubing, syringe and extra stopcock. We recommend you replace the transducer 
dome and tubing with sterile ones.
8
Label the transducer with the calibration factor shown in the 
Cal. Factor
 field in the pressure’s 
setup menu.
9
Reconnect the patient and start measuring again.
Troubleshooting the Pressure Calibration
The status line lists the probable cause of an unsuccessful calibration.
Calculating Cerebral Perfusion
The monitor can calculate the difference between mean arterial pressure and the intracranial pressure. 
The difference is labeled CPP.
1
In the 
Main
 
Setup 
menu, select 
Measurements
.
2
In the 
Setup
 
CPP 
menu, select
 ABP
ART
Ao
BAP
 or 
FAP 
as the arterial pressure source.
Calculating Pulse Pressure Variation 
Note: The derived measurement Pulse Pressure Variation is not available in the USA.
Pulse Pressure Variation (PPV) is calculated from beat-to-beat arterial pressure values. Pulse pressure is 
the difference between the systolic and the diastolic pressure values for a single beat. The average 
variation in pulse pressure is calculated over periods of 32 seconds.
WARNING
• This monitor can calculate PPV from beat-to-beat values of any arterial pulsatile pressure. The 
circumstances under which the calculation of a PPV value is clinically meaningful, appropriate and 
reliable must be determined by a physician. 
Message
Corrective Action
unable to calibrate - 
equipment malfunction
Contact your service department. The pressure hardware is faulty.
unable to calibrate - 
out of range
Make sure that you have selected the value for 
Cal. Press
 that 
you are applying to the transducer, and repeat the calibration.
unable to calibrate - no 
transducer
Make sure that the transducer is connected and try again.
unable to calibrate - 
unstable signal
Make sure there are no disturbances to the transducer, and repeat 
the calibration.
unable to calibrate - 
perform zero first
No valid zero. Zero the transducer.