Emerson 2200S User Manual

Page of 158
92
Micro Motion
®
 Model 2200S Transmitters
Operating the Transmitter
8.7.2
Controlling totalizers and inventories
Table 8-1 shows all of the totalizer and inventory functions and which configuration tools you can use 
to control them.
With the display
Table 8-2 shows how you can control the totalizers and inventories with the display.
Table 8-1
Totalizer and inventory control methods
Function name
Display
Communicator
ProLink II
Start/stop all totalizers and inventories
Yes
(1)
Yes
Yes
Reset mass totalizer value only
Yes
Yes
Yes
Reset volume totalizer only
Yes
Yes
Yes
Simultaneously reset all totalizer values
Yes
Yes
Yes
Simultaneously reset all inventory values
No
No
Yes
(2)
(2) If enabled in the ProLink II Preferences window.
Reset mass inventory value only
No
No
Yes
Reset volume inventory value only
No
No
Yes
Table 8-2
Totalizer and inventory control with the display
To accomplish this
Press this sequence of buttons
Stop all totalizers and 
inventories
(1)
(1) This feature may be enabled or disabled. See Section 5.2.4.
• Scroll until a totalizer value appears (the word 
TOTAL
 appears in the lower left corner of 
the display). It does not matter whether the total is mass or volume.
• Select.
• Scroll until 
STOP
 appears beneath the current totalizer value.
• Select (
YES
 alternates with 
STOP
).
• Select (all totalizers and inventories stop).
• Scroll to 
EXIT
.
Start all totalizers and 
inventories
• Scroll until a totalizer value appears (the word 
TOTAL
 appears in the lower left corner of 
the display). It does not matter whether the total is mass or volume.
• Select.
• Scroll until 
START
 appears beneath the current totalizer value.
• Select (
YES
 alternates with 
START
).
• Select (all totalizers and inventories start).
• Scroll to 
EXIT
.
• Select.
Reset mass totalizer
• Scroll until the mass totalizer value appears.
• Select.
• Scroll until 
RESET
 appears beneath the current totalizer value.
• Select (
YES
 alternates with 
RESET
).
• Select (mass totalizer resets).
• Scroll to 
EXIT
.
• Select.
Reset volume (liquid or 
gas) totalizer
• Scroll until the volume totalizer value appears.
• Select.
• Scroll until 
RESET
 appears beneath the current totalizer value.
• Select (
YES
 alternates with 
RESET
).
• Select (volume totalizer resets).
• Scroll to 
EXIT
.
• Select.