Emerson 2200S User Manual

Page of 158
Configuration and Use Manual
101
Measurement Performance
Op
eratio
n
App
e
ndi
ces
Mai
n
tenan
c
e and
 T
roub
le
sh
oo
tin
g
Com
m
is
si
on
in
g
9.5
Performing temperature calibration
Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope 
calibration. The entire procedure must be completed without interruption.
Note: Before performing the calibration, record your current calibration parameters. You can do this 
by saving the current configuration to a file on the PC. If the calibration fails, restore the known 
values.
You can calibrate for temperature with ProLink II software. See Figure 9-3.
Figure 9-3
Temperature calibration – ProLink II
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves 
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress 
light turns green
Calibration in Progress 
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves 
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress 
light turns green
Calibration in Progress 
light turns red
Close
Done
ProLink Menu >
     Calibration >
          Temp offset cal
ProLink Menu >
     Calibration >
          Temp slope cal