9.5
Performing temperature calibration
Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope
calibration. The entire procedure must be completed without interruption.
calibration. The entire procedure must be completed without interruption.
Note: Before performing the calibration, record your current calibration parameters. You can do this
by saving the current configuration to a file on the PC. If the calibration fails, restore the known
values.
by saving the current configuration to a file on the PC. If the calibration fails, restore the known
values.
You can calibrate for temperature with ProLink II software. See Figure 9-3.
Figure 9-3
Temperature calibration – ProLink II
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal
Calibration >
Temp slope cal