Appendix A
Default Values and Ranges
A.1
Overview
This appendix provides information on the default values for most transmitter parameters. Where 
appropriate, valid ranges are also defined. 
These default values represent the transmitter configuration after a master reset. Depending on how 
the transmitter was ordered, certain values may have been configured at the factory. 
A.2
Most frequently used defaults and ranges
The table below contains the default values and ranges for the most frequently used transmitter 
settings. 
Table A-1
Transmitter default values and ranges
Type
Setting
Default
Range
Comments
Flow
Flow direction
Forward
Flow damping
0.64 sec
0.0 – 60.0 sec
User-entered value is 
corrected to nearest lower 
value in list of preset values. 
For gas applications, 
Micro Motion recommends a 
minimum value of 2.56.
Mass flow units
g/s
Mass flow cutoff
0.0 g/s
Recommended setting is 5% 
of the sensor’s rated 
maximum flowrate.
Volume flow type
Liquid
Volume flow units
L/s
Volume flow cutoff
0.0 L/s
0.0 – x L/s
x is obtained by multiplying 
the flow calibration factor by 
0.2, using units of L/s.
GSV cutoff
0.0
Meter factors
Mass factor
1.00000
Density factor
1.00000
Volume factor
1.00000