Appendix A
Default Values and Ranges
A.1
Overview
This appendix provides information on the default values for most transmitter parameters. Where
appropriate, valid ranges are also defined.
appropriate, valid ranges are also defined.
These default values represent the transmitter configuration after a master reset. Depending on how
the transmitter was ordered, certain values may have been configured at the factory.
the transmitter was ordered, certain values may have been configured at the factory.
A.2
Most frequently used defaults and ranges
The table below contains the default values and ranges for the most frequently used transmitter
settings.
settings.
Table A-1
Transmitter default values and ranges
Type
Setting
Default
Range
Comments
Flow
Flow direction
Forward
Flow damping
0.64 sec
0.0 – 60.0 sec
User-entered value is
corrected to nearest lower
value in list of preset values.
For gas applications,
Micro Motion recommends a
minimum value of 2.56.
corrected to nearest lower
value in list of preset values.
For gas applications,
Micro Motion recommends a
minimum value of 2.56.
Mass flow units
g/s
Mass flow cutoff
0.0 g/s
Recommended setting is 5%
of the sensor’s rated
maximum flowrate.
of the sensor’s rated
maximum flowrate.
Volume flow type
Liquid
Volume flow units
L/s
Volume flow cutoff
0.0 L/s
0.0 – x L/s
x is obtained by multiplying
the flow calibration factor by
0.2, using units of L/s.
the flow calibration factor by
0.2, using units of L/s.
GSV cutoff
0.0
Meter factors
Mass factor
1.00000
Density factor
1.00000
Volume factor
1.00000