Emerson 2200S User Manual

Page of 158
32
Micro Motion
®
 Model 2200S Transmitters
Configuring Process Measurement
Configuration
To characterize the flowmeter using the Communicator:
1. Select 
Detailed Setup > Charize Sensor
.
2. In the Sensor Selection menu, select 
Other
.
3. In the Flow menu, set the 
FlowCal
 parameter.
4. In the Density menu, set the remainder of the parameters listed in Table 4-1.
To characterize the flowmeter using ProLink II:
1. Click 
ProLink > Configuration.
2. On the Device panel, set 
Sensor Type
 to 
Curved
.
3. On the Flow panel, set the 
Flow Cal
 parameter.
4. On the Density panel, set the remainder of the parameters listed in Table 4-1.
4.3
Configuring general flow parameters
The general flow parameters include:
Flow Direction
Flow Damping
4.3.1
Flow direction
The Flow Direction parameter controls how the transmitter reports flow rate and how flow is added to 
or subtracted from the totalizers, under conditions of forward flow, reverse flow, or zero flow.
Forward (positive) flow moves in the direction of the arrow on the sensor.
Reverse (negative) flow moves in the direction opposite of the arrow on the sensor.
Table 4-1
Characterization parameters
Parameter
Description
Sample sensor tag
K1
Tube period when sensor is filled with air
K2
Tube period when sensor is filled with water
FD
Density correction factor for high flow rates
D1
Density of air for K1
D2
Density of water for K2
TC
Temperature coefficient to compensate for the 
effect of temperature on the density 
measurement
Flowcal
Flow calibration factor to define the 
relationship between sensor data and mass 
flow rate and compensate for the effect of 
temperature on the mass measurement
19.0005.13
0.0010
0.9980
12502.000
14282.000
4.44000
310
12500142864.44