Emerson Process Management 2400S User Manual

Page of 208
88
Micro Motion
®
 Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
There are two versions of the meter verification application: the original version and Micro Motion 
Smart Meter Verification. Table 10-1 lists requirements for each version. Table 10-2 provides a 
comparison of the two versions.
Note: If you are running an older version of ProLink II or the EDD, you will not be able to access the 
additional features in Smart Meter Verification. If you are running an updated version of ProLink II 
or the EDD with the original version of meter verification, the meter verification procedures will be 
slightly different from the procedures shown here.
Table 10-1
Version requirements for meter verification application 
Requirement type
Meter verification application
Original version
Smart Meter Verification
Transmitter
v1.0
v1.4
ProLink II requirements
v2.5
v2.9
EDD requirements
2400SDP_pdmrev1_00 folder
2400SDP_pdmrev1_40 folder
Table 10-2
Comparison of meter verification features and functions: original version vs. Smart Meter 
Verification 
Feature or function
Meter verification application
Original version
Smart Meter Verification
Process interruption
No need to halt flow
No need to halt flow
Measurement interruption
Three minutes. Outputs go to: 
• Last Measured Value
• Configured Fault Value
User option: 
• Continue Measurement. Measurement is 
not interrupted. Test requires 
approximately 90 seconds.
• Last Measured Value. Outputs fixed and 
measurement interrupted for 
approximately 140 seconds.
• Configured Fault Value Outputs fixed and 
measurement interrupted for 
approximately 140 seconds.
Result storage
Test results stored only for tests run with 
ProLink II, and stored on the PC
Twenty most recent results stored on the 
transmitter, independent of tool used to 
perform the procedure. For tests run with 
ProLink II, additional result data stored on 
PC.
Result data on display
Pass/Fail/Abort for current test
For all results stored on transmitter: 
• Pass/Fail/Abort
• Abort code (if relevant)
• Stiffness of the right and left pickoffs