Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 4: Configuration
3-9000-765 Rev H
May 2015
Field Setup Wizard
 77
Display units
The Meter volume units displayed are either U.S. Customary or Metric. To modify the Display 
Units, configure the Meter units system in the Field Setup Wizard — General Page. 
U.S. Customary volume unit selections are:
Barrels
Gallons
Metric volume unit selections are:
Cubic meters
Cubic liters
Display units preceded by a plus or minus sign indicate forward and reverse flow 
direction.
The Local Display Flow rate time units are modifiable by selecting the drop-down 
arrow and clicking the time unit in the list box.
Valid flow rate time units selections are:
second
minute
hour
day
Scroll delay
The Scroll Delay is the time interval for the selected display items to be shown on the Local 
Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow to 
increase or decrease the length of time an item displays.
1.
Select 
Finish
 to write the configuration settings to the meter.
2.
Save the meter configuration file, collect a Maintenance log and Waveforms to 
document the “As Left” settings.