John Bunn O Premier Oxygen Concentrator
trator’s AC connector into a properly grounded power 
outlet.
turning the Concentrator On
1.  Press the power switch to start the concentrator. There 
will be a brief delay before the concentrator starts. The 
alarm system emits a tone of one second duration, the 
word HELLO appears on the hour meter LED panel, and 
the green power indicator light illuminates.
ATTENTION: Every time the concentrator starts, the alarm 
system performs a self test, and emits a brief tone. If 
the alarm does anything other than emit a brief tone, 
see Section 6, Troubleshooting Guide, and follow the 
appropriate steps.
2.  Turn the flowmeter knob to the setting prescribed by 
your physician or therapist.
Note: To properly read the flowmeter, locate the prescribed
flowrate line on the flowmeter.
3.  Turn the flowmeter knob until the flowmeter ball rises 
to the appropriate line.
4.  Center the flowmeter ball on the L/min line prescribed.
s
Damage: If the flowrate on the flowmeter ever falls below 
0.5L/min, check tubing or accessories for blocked or 
kinked tubing or for a defective humidifier bottle.
If using a humidifier bottle, the oxygen comes from the 
humidifier bottle outlet.
5.  Connect the oxygen tubing barb connector securely to 
the humidifier outlet; the other end of the oxygen tub-
ing is used by the patient.
Note: Oxygenation times and the flowrate ranges are
established and prescribed by your physician.
turning the Concentrator Off
Turn off the power switch and unplug the concentrator’s 
AC connector from the power outlet.