TRANSDUCER ASSEMBLY SECURITY SEALS
Section 3: Electrical installation
Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014
3-9000-759 Rev D
3.7.3
Transducer assembly security seals
Use the following instructions and 
 to install the security seal wire on the transducer 
assembly.
1.
Rotate the transducer cable nut (Item A) clockwise, compressing the seal on the 
transducer cable connector.
2.
Insert a security seal wire into one of the two holes in the transducer cable connector 
(Item B) and feed through one of the two holes in the transducer cable nut (Item A). 
Note
: Choose holes that minimize counterclockwise rotation of the transducer cable 
nut when the security wire (Item C) is taut.
3.
Adjust the security wire, removing all slack and thread into the lead seal.
4.
Cut wire ends to remove excess wire.
Figure 3-14  Transducer assembly security seal
A.
B.
C.
A.  Transducer cable nut
B.  Transducer cable connector
C.  Security wire seal