If the batteries are dead and you cannot replace them, 
switch to Continuous Flow mode by pulling the CF switch out.
The unit will deliver 2 liters per minute of flow and cannot be 
adjusted.
Continuous Flow mode does not require batteries.
See the “Changing the Batteries” (pg 14) section in this 
manual for more information.
6. To stop continuous, push the CF switch 
in and hold the power button until the 
power is off. The lights will illuminate 
and count down from 5-1 and then
shut off.
If the SmartDose Liquid Oxygen Portable is in Continuous 
Flow mode, a small amount of oxygen will still be delivered. 
The portable must be powered off and the CF switch pushed 
in to completely turn off the flow of oxygen.
If using rechargeable batteries, it is recommended to use 
>1800 mAh Ni-MH (Nickel Metal Hydride) batteries. Do NOT 
mix battery types.
13
12
   
 
 
 
 
 
   
 
   
 
   
 
 
 
 
NOTE
Starting Oxygen Delivery from the SmartDose Liquid Oxygen
Portable
To receive oxygen from your SmartDose Liquid Oxygen Portable, follow
these steps.
1. Push the cannula breathing tube 
firmly onto the oxygen outlet 
port. Adjust the cannula breathing 
tube to the proper position on your 
nose so that you will be able to 
breathe the oxygen comfortably.
2. Turn on the SmartDose Liquid Oxygen Portable by pressing and 
3.
holding the Power Button until all of the numerical LEDs 
(1 through 5) light up.
Be sure that the CF (Continuous 
Flow/SmartDose) switch is pushed 
in as shown, confirming that your 
Liquid Oxygen Portable is in the 
SmartDose mode.
4. Gaseous oxygen is now flowing from your SmartDose Portable, 
and with each breath you should receive your set dose of oxygen 
and the set Dose Indicator Light will illuminate with each breath.
In order for the SmartDose Portable to determine resting breath 
rate, the patient must rest 1-2 minutes before beginning activity.
5. Keep an eye on your SmartDose Liquid Oxygen Portable’s battery 
power.
• The Battery Indicator Light only monitors battery power and 
does not indicate whether there is liquid oxygen in the reservoir.
• When the battery indicator lights up red with each breath, it is 
time to replace the batteries.
• The battery indicator will remain lit red (not blinking with each 
breath) if the batteries are totally depleted.
 
   
 
   
 
   
 
 
 
 
   
 
   
 
   
 
 
 
 
NOTE
 
   
 
   
 
   
 
 
 
 
   
 
   
 
   
 
 
 
 
NOTE
Activation Button
Set your prescribed setting (1 to 5) 
by pressing the Activation Button 
until the correct setting is illuminated.
There is no need to change your setting with the 
Activation Button.