Chapter 8. Servicing the PT878
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
201
8.12.2b Installing New Software (cont.)
5.
In the Hyperterminal window, pull down the Transfer menu and click
Send File. A window opens similar to Figure 135 below.
Send File. A window opens similar to Figure 135 below.
Figure 135: The Send File Window in the Transfer Menu
6.
Click on the replacement software (designated by a .cod extension)
from the folder where it has been stored.
from the folder where it has been stored.
7.
From the Protocol drop-down menu, select Xmodem.
8.
Click Send. For a successful transfer, the PC window appears similar to
Figure 136 on page 201, while the PT878 screen displays the program
ID, size, load address and a count of blocks being loaded.
Figure 136 on page 201, while the PT878 screen displays the program
ID, size, load address and a count of blocks being loaded.
Figure 136: The File Transmission Window