Konica Minolta PULSOX-300i User Manual

Page of 40
8
Foreword
Notes on Use
•  This instrument should be used under the following operat-
ing conditions.
  Temperature of 0 to 40°C (32 to 104°F), relative humidity 
of 30 to 85%, with no condensation.
  Atmospheric pressure of 700 to 1060 hPa (altitude of 
400 to 3000m).
•  This is a precision instrument.  To avoid the possibility of it 
being damaged, the instrument should not be dropped nor 
should heavy objects be placed on top of it.
•  Do not expose this instrument to rain and water.
•  This instrument has no alarm function. Do not use this in-
strument if an alarm function is  necessary.
•  This instrument is designed for use on adults. When using 
the instrument on infants or babies, please note that the de-
signed measurement accuracy cannot be guaranteed.
•  Do not use the instrument on a point where there is exces-
sive vibration. In addition, to avoid excessive impact on the 
instrument, it should be handled gently. Failure to observe 
this may cause breakdown.
•  A maximum of 300 hours of measurement data can be stored, 
or the maximum storable number of data files is 399 files. If 
measurements are continued after the maximum number of 
storable data has been reached, the new data will overwrite 
the oldest data. To keep data, be sure to download it before 
the memory becomes full.
•  Accurate measurement data may not be obtained in the fol-
lowing cases. 
  When body movement is excessive
  When the probe is not attached properly
  When blood circulation at the measuring point is poor (due to 
pressure on arm/finger, peripheral circulatory insufficiency)
  When ambient light (e.g. panel light, fluorescent light, in-
frared heat lamp, direct sunlight) is too strong