Konica Minolta PULSOX-300i User Manual

Page of 40
19
Measurement
2) Reading the display
When the instrument is turned ON, all the elements of the LCD 
will light up. 
The pulse level, pulse rate and SpO
2
 value are displayed 
during measurement.
Pulse rate (P.R.)
If an error occurs, an 
error message mainly 
a b o u t   P. R .   w i l l   b e 
displayed.
Battery mark
I n d i c a t e s   t h e 
remaining battery 
capacity in three 
steps.
Pulse Level Meter
Oxygen  saturation 
(SpO
2
) value
If  an  error  occurs,  an 
error  message  mainly 
a b o u t   S p O
2
  w i l l   b e 
displayed.
Indicate the pulse level.
  For stable measurement, make 
sure that the pulse level meter 
lights up two or more segments 
by adjusting the measuring point 
or rubbing or warming it up to 
improve blood circulation.