ResMed AUTO 25 User Manual

Page of 22
14
Displays error message: 
Check tube!! Key if done
Displays error message: 
Exxxx Call Service
 (where 
xxxx
 defines an error)
Displays error message: 
HIGH LEAK!!! Adjust Mask
The following message is displayed on the LCD after you try to update settings or copy data to 
the Data Card: 
Card Error Remove Card
 
The following message is displayed on the LCD after you try to update the settings using the 
Data Card: 
Settings Invalid Remove Card
The following message is displayed on the LCD after you try to update the settings using the 
Data Card: 
Settings Error Remove Card
The following message is NOT displayed on the LCD after you try to update the settings using 
the Data Card: 
Settings Success Remove Card
Problem/Possible Cause
Solution
The air tubing is loose or blocked.
Check that the air tubing is connected securely to your mask 
and the air outlet. Press the Start/Stop key to restart the 
device. If this does not clear the message, disconnect the 
power cord and then reconnect it to restart the device.
Component failure.
Record error number and contact your ResMed service center.
You have experienced excessively 
high leak levels for more than 20 
seconds.
Check that your air tubing is connected properly. 
Adjust headgear.
Data Card is not inserted 
correctly.
Ensure that the Data Card is inserted with the arrow facing up 
as far as it can go.
You may have removed the Data 
Card before settings were copied 
to the VPAP device.
Reinsert the Data Card and wait for the Settings Success 
Remove Card
 or Copy Complete Remove Card message 
to appear on the LCD.
The identification details on the 
Data Card do not match the 
details on your device.
Contact your clinician/service provider immediately.
There is a data error on the Data 
Card.
Contact your clinician/service provider immediately.
The settings were not updated.
Contact your clinician/service provider immediately.