Invacare IRC5PO2 User Manual
SECTION 5—CABINET
23
SECTION 5—CABINET
Removing Cabinet
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
before servicing.
NOTE: For this procedure, refer to FIGURE 5.1.
1. Unplug unit.
2. Remove the four mounting screws that secure cabinet assembly to the base assembly.
3. Lift the cabinet straight up.
NOTE: When required, vacuum inside of the cabinet and exposed foam insulation.
4. To re‐install cabinet, reverse STEPS 2‐3.
L
FIGURE 5.1 Removing Cabinet
Cabinet Assembly
Mounting
Screws
Base Assembly
Mounting
Screws