Tektronix WFM 601E User Manual

Page of 216
Getting Started
1–2
WFM 601A, WFM 601E & WFM 601M User Manual
 SMPTE RP-165 standard EDH indicator for presence
 Embedded Audio indentification
 Analog audio Lissajous display on the WFM 601A
 Source signal level meter and cable-length readout on the WFM 601E and
the WFM 601M
 Lightning and Vector displays
 Reclocked Serial Component Digital output following A/B switching
 Video Reference: Internal Serial Component signal (follows A/B switching)
or external composite
An expanded feature set is available through the menus. You select menu items
with multi-use bezel buttons and knobs. When you select a menu item, such as
Voltage Cursors, on-screen labels show the current function of the bezel buttons
and knobs.
The waveform monitors provide an internal calibrator signal to set both vertical
and horizontal gain. The calibrator signal is a 700 mV, 100 kHz signal. Press the
CONFIG button and use the bezel knob to select the Calibrate menu. Press the
CAL SIG button to turn on the calibrator signal.
Accessories
The waveform monitor is shipped with several standard accessories. These
standard accessories and any optional accessories are listed here.
The following accessory items are included with this product:
1
User Manual, this manual (071-0103–XX)
1
Power Cord: United States and Japan only (161–0216–XX)
1
Replacement Fuse Cartridge: 3AG, 2A, 250 V, fast-blow (159–0021–00)
3
Replacement Graticule Light Bulbs (150–0168–00)
3
Replacement Air Filters for Fan (378–0415–00)
2
75 
W High-frequency, End-line Terminations: 26 dB to 300 MHz
(011–0163–00)
1
Smoke Grey CRT Filter, installed on instrument (378–0258–00)
Menus
Calibrator
Standard Accessories