Tektronix WFM 601A User Manual

Page of 216
Display Commands
WFM 601A, WFM 601E & WFM 601M User Manual
C–29
sponds to the BLACK LEVEL adjustment available when the CRT MENU
button is pressed in the Picture display.
DISPlay:PIXB  <blackness range>
DISPlay[:PIXB]?
<blackness range>
displ:pixb 100
display:pixb?
:DISPLAY:PIXBLACKNESS 100
DISPlay:TRACEROT  <tracerotate>
Allows adjustment of trace rotation. The argument <tracerotate> is an integer
from –2047 to +2047. The median value ( 0 ) represents a trace rotation angle of
approximately 0 degrees. Each extreme value rotates the trace by about 15
degrees.
DISPlay:TRACEROTate  <tracerotate> 
DISPlay[:TRACEROTate]?
<tracerotate>
disp:tracerot –100
display:tracerot?
:DISPLAY:TRACEROTATE –100
DISPlay:SCALEillum  <scale>
Adjusts the brightness level of the permanent scale on the CRT. The brightest
value for the argument <scale> is 0 and the lowest is –2047. The best value
depends on the brightness of the environment.
DISPlay:SCALEillum  <scale> 
DISPlay[:SCALEillum]?
Syntax
Arguments
Examples
Returns
Syntax
Arguments
Examples
Returns
Syntax