Anritsu MG3641A User Manual

Page of 212
SECTION 4   OPERATING INSTRUCTIONS
4-44
(3) Execution using the panel key
When the "Exe" [F4] key on the "Trigger Program" key
is pressed, the trigger program is executed.  This method
can be used to check the trigger function.
The trigger program can be halted by pressing a key other than the [Preset] or [Local] key on the panel.
4.11.4 Checking the contents of the trigger program
The contents of the trigger program can be checked by opening the "Trigger Program" menu.  In the trigger program,
one step is displayed as one line.  If the trigger program consists of more than one page, a page number is displayed
in the menu title.  In this case, press the [More] key to advance to the next page.
Trigger Program
SYST BELL OFF
FREQ SWE TYPE 0
FREQ SWE PATT LOG
FREQ SWE STAR 1 MHz
FREQ SWE STOP 100 MHz
Clr
Rtn
Executing the trigger program
Exe
Entr
Trigger Program (1/2)
SYST BELL OFF
FREQ SWE TYPE 0
FREQ SWE PATT LOG
FREQ SWE STAR 1 MHz
FREQ SWE STOP 100 MHz
Clr
Rtn
Exe
Entr
[More] キー
Trigger Program (2/2)
FREQ SWE TIME 100 ms
FREQ SWE RUN 
Clr
Rtn
Exe
Entr