A&D EJ-3000 User Manual

Page of 42
 33
 
2.   CC 0    is displayed. 
 
3. If necessary to change the weight value to check, press the   SAMPLE   key and 
change the weight value using the following keys. 
 
  SAMPLE    key 
To select the digit blinking to change. 
  RE-ZERO    key  To set the value of the digit selected.   
  PRINT    key 
To store the value and return to step 2. 
 
4. At step 2, press the    PRINT    key. The zero point 
is weighed and the weighed value is displayed for a 
few seconds. 
 
 
 
 
 
 
 
 
 
5. Place the displayed weight on the pan and press 
the  
 PRINT  key to weigh it. The weighed value is 
displayed for a few seconds. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
6.   end     appears. 
 
 
7.   glp      is displayed and calibration test report is 
output. 
 
 
8.   end     appears again. Remove the weight and 
press the  MODE  key to return to the weighing 
mode. 
 
 
 
  CC  0 
     0.0 
  CC  0 
The weighed data
is shown with “g”.
With nothing 
on the pan 
  2000.0 
  2000.0 
  2000.0 
Displayed 
weight 
 end 
 end 
The weighed data
is shown with “g”.
Returns to   
the weighing mode 
  glp 
GLP output