A&D SW-6KS User Manual

Page of 46
37
 
 
Command examples    (“_” indicates “Space” (20H).) 
The examples below are for the function setting “
aCk 1
 “ (Reply to commands). 
The address ## = 23. 
‰
 Request a weight data. 
Command 
@ 2 3  Q C
R
  L
F
 
 
Reply 
@  2 3  S  T  ,  + 0 0 1 2 3 4 5 _ k g  C
R
  L
F
 
Stable data 
 
@  2 3  U  S  ,  + 0 0 0 7 . 8 9 0 _ k g  C
R
  L
F
 
Unstable data
 
@  2 3  O  L  ,  + 9 9 9 9 . 9 9 9 _ k g  C
R
  L
F
 
“E” display 
 
‰
 Zero the scale. (No reply for the function setting “
aCk 0
 “.) 
Command 
@ 2 3  Z C
R
  L
F
 
 
 
Reply 
@ 2 3  Z C
R
  L
F
 
 
The scale is in a condition that zero operation is possible.
 
‰
 Send the current LO limit value. 
Command 
@ 2 3 ? L 1 C
R
L
F
 
Reply 
@ 2 3 L 1  ,  + 0 0 0 2 0 0 C
R
L
F