 11
 
9. The balance re-calculates the unit weight while the ACAI indicator is blinking. Do not 
touch the balance or samples on the pan until the ACAI indicator turns off. 
 
10. Counting accuracy is improved when the ACAI indicator turns off. Each time the 
above operation is performed, a more accurate unit weight will be obtained. There is 
no definite upper limit of ACAI range for the number of samples exceeding 100. Try 
to add the similar number of samples as displayed. 
 
 
5-7. Percent mode (%) 
Displays the weighing value in percentage compared to the reference (100%) weight.     
 
Selecting the percent mode 
1. Press the  UNITS  key to select %. (%:percent) 
  
 
 
 
 
   
 
 
Storing the reference (100%) weight 
2. 
Press the  SAMPLE  key to enter the 
reference weight storing mode. 
 
 
3. Press the    RE-ZERO    key to display 100 0%.  
 
 
 
 
4. Place the sample to be set as the reference 
weight on the pan. 
 
 
 
 
 
 
5. Press the   PRINT   key to store the reference 
weight. Remove the sample. 
 
 
 
 
 
Reading the percentage 
6. 
Place the object to be compared to the 
reference weight on the pan. The displayed 
percentage is based on 100% of the reference 
weight. 
 
 
 
 
To store
     0.0 
%
  100 - 
%
   100.0 
%
    75.8 
%
  100 0 
%
ZERO   
Weighing pan
Sample corresponding
to the 100% weight 
Object to be
compared