A&D EJ-440 User Manual

Page of 42
 29
 
2. Place a sample on the upper pan.   
 
 
3. Wait for the STABLE indicator to be displayed 
and press the  SAMPLE  key to store the 
weight in air. 
 
 
 
4.  The display shows that 
 LO 
 blinks and the 
balance starts to measure weight in water. 
 
 
5. Place the sample on the lower pan in water. 
‰
 Adjust the amount of water so that the sample 
is about 10 mm below water surface. 
 
6. Wait for the STABLE indicator to be displayed 
and press the  SAMPLE  key. Then the 
balance reads the weight in water and shows 
the density (specific gravity) of the sample. 
 
 
 
 
7. To continue the specific gravity measurement, 
press the  SAMPLE  key again. To exit this 
measurement, press the  UNITS  key. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   20.00 
SG 
HI 
   20.00 
SG 
LO 
Measuring 
weight in water. 
   17.51 
SG 
LO 
To weigh in water.
To weigh in air.
 
The display shows
the density.
 
   
8
8.01 
SG