5. OPERATION 
 
5-1. Turn the power ON and OFF 
 
1. Press the    ON:OFF    key to turn the power ON. 
 
 
 
 
 
 
 
 
All of the symbols are displayed as shown above. 
(About units: Only the units available will be displayed.) 
 
The display turns off except for a weighing unit and the decimal point. 
The balance waits for the weighing data to become stable, and zero will be 
displayed with the ZERO indicator (power-on zero). 
 
The range for power-on zero is within ±10% of the weighing capacity around the 
calibrated zero point. 
If the power is switched ON while there is a load beyond this range, the balance will 
be tared to zero and the NET indicator and the ZERO indicator turn on. 
 
2. Press the    ON:OFF    key again, and the power will be switched OFF. 
 
‰
  Auto-power off function 
  It is possible to have the power automatically switched OFF, if zero is 
displayed for approximately 5 minutes. See “8-5. Function list” and 
set the function “
poff
”. 
 
 
5-2. LCD backlight 
 
The LCD backlight will turn on when the weight value changes more than 4d (4 x min.   
display division) or any key operation is done. When the weight data becomes and 
stays stable for some moment, the backlight will automatically turn off. There is also a 
setting that the backlight is always on or off. For details, see the function setting “ltUp” 
of “Function list”. 
 
 
 
 
 
 
 
 
 
 
 
 
STABLE
NET
ZERO
Weighing 
units