7.
7.
7.
7.    Tare
Tare
Tare
Tare    
  The function is used to display a net value with the container weight subtracted from 
the total weight, if you place an object into a container to weigh it. 
  Using a serial interface such as the RS-232C, you can do this from the external 
equipment. 
 
Caution 
  When turning the display off with 
Cf02 1
Cf02 1
Cf02 1
Cf02 1, 2222 or 3333, the tare data is cleared. 
 
  When turning the power off, the tare data is cleared. 
 
7.1.1. Weighing 
Tare 
Operation 
Place the tare on the weighing unit. Press the 
TARE
 key to store the tare 
weight after the STABLE mark turns on. The display changes to net. 
 
Caution 
  When displaying a negative gross value, tare can not be used (normally). 
 
  To enable tare at zero or a negative gross value, select a 
Cf 04
Cf 04
Cf 04
Cf 04 parameter. 
 
7.1.2. 
Digital Input (Preset Tare) 
Caution 
  The input value is rounded off to the unit of division(weighing interval). 
 
  In the case of 
Cf02 1
Cf02 1
Cf02 1
Cf02 1, 2222 or 3333 (power-on zero), the displayed value will be 
zeroed when turning the display on. 
 
  When using a multi-interval scale, the usable input range is the first range. 
 
  In the case of 
Cf09 1
Cf09 1
Cf09 1
Cf09 1 (To inhibit preset tare), preset tare can not be used. 
 
  Preset tare is stored as one of the Code Memory data. Refer “9. Code 
Memory”. 
 
  Preset tare value can be set via the serial interface. 
 
Step 1 
Press the PT key to display the stored tare value. 
 
When tare is cleared or is not used, the value is zero. 
Step 2 
Enter a new tare value using the 
0
 - 
9
 keys. 
ENTER
 key 
The key to store a new tare value. 
 
The net is displayed. 
ESC
 key 
The key to return to weight display without changing the stored 
value. 
 
7.1.3. Clearing 
Tare 
 
  When pressing the 
TARE
 key while gross is zero, tare is cleared and gross is 
displayed. 
 
  When zeroing with the 
ZERO
 key, tare is cleared. 
 
  
7. 