  There are 3 type of setting values for each comparison. 
(1) Set the limit value (upper and lower limit / HiHI, Hi, Lo, LoLo limit). 
(2) Set the Target value and an acceptable tolerance (upper and lower) in weight. 
  The limit value is calculated automatically. 
(3) Set the Target value and an acceptable tolerance (upper and lower) in 
percentage of the target weight. The limit value is calculated automatically. 
Example.  Target = 50kg,  Upper limit = 51kg,  Lower limit = 48kg 
(1) Hi (Upper limit): 51 (kg), Lo (Lower limit): 48 (kg) 
(2) TG (Target):    50 (kg), Hi (Upper acceptable tolerance): 1 (kg),   
 
Lo (Lower acceptable tolerance): 2 (kg) not a negative value 
(3) TG (Target): 50 (kg), Hi (Upper acceptable tolerance): 2 (% of Target), Lo 
(Lower acceptable tolerance): 4 (% of Target) not a negative value 
 
 
10.1.1.  Condition formula for Comparison 
Comparison is performed based on the following formula. 
Upper/lower comparison 
Judge Condition 
Formula Display 
Output
HI 
Upper limit (Hi limit) value 
¢
 Displayed value 
HI 
HI 
OK 
Lower limit value 
*
 Displayed value 
*
 Upper limit value 
OK 
OK 
LO Displayed 
value 
¢
 Lower limit (Lo limit) value 
LO 
LO 
5-stage comparison 
Judge Condition 
Formula Display 
Output
HiHi 
HiHi limit value 
¢
 Displayed value 
HI,  2  
HI 
HI 
Upper limit (Hi limit) value 
¢
 Displayed value 
HI 
HI, OK
OK 
Lower limit value 
*
 Displayed value 
*
 Upper limit value 
OK 
OK 
LO Displayed 
value 
¢
 Lower limit (Lo limit) value 
LO 
LO, OK
LoLo Displayed 
value 
¢
 LoLo limit value 
LO,    3   
LO 
 
 
  The decimal point is not considered. Example: If the upper limit value is 10.0, enter 100. 
 
  These parameters are stored in non-volatile memory, and are not lost even if the 
AC power is disconnected or cut off (except code 0 memory). 
 
  When the displayed value becomes an overload (positive over), HI (over) is output. 
When the displayed value becomes an under load (negative over), LO (under) is 
output. 
 
  This function compares the Hi / HiHi limit value first. 
 
  This function does not check the relationship between the upper and lower limit 
values. 
 
 
 
10.1. Weight Check Mode