AD-4405 Weighing Indicator
 
Description of "Command Mode" (For Serial Interface) 
Function 
f40 5 
Operation 
The serial data can be output by the serial interface command only. Data 
conflicts or accidental data output by mis-operation can be avoided when 
using multiple instruments with the RS-422/485 interface.   
Description of "Delay for continual data" 
Function 
f38, f42 
Operation 
This function can be used in the "Auto print mode" and "Manual print mode". 
 
 
When using a non-buffered printer, set to f42 3 and f43 1 (or f38 1). 
 
15.2.
15.2.
15.2.
15.2.    Data 
Number
Data Number
Data Number
Data Number    
The data number increments automatically with each serial data output or print out 
respectively from 1 to 99999 (the next of 99999 is 1). The starting number can be set 
using the keys. 
Setting the Serial Data Number 
Press the 
CODE(1)
 key while pressing the 
PRINT
 key at the weight display, the display 
changes to data r then r xxx. “xxx” is the next data number. Use the 
0
 - 
9
 keys to enter 
the number, then press the 
ENTER
 key to store the number and return to the weight 
display. 
Setting the Print Data Number 
Press the 
SAMPLE(2)
 key while pressing the 
PRINT
 key at the weight display, the display 
changes to data p then p xxx. “xxx” is the next data number. Use the 
0
 - 
9
 keys to enter 
the number, then press the 
ENTER
 key to store the number and return to the weight 
display. 
Note 
1. At the number setting, the following keys are available. 
 
 
+/- key 
Sets the number to 1. 
 
 
ZERO
 key  Sets the number to 0. 
 
 
ESC 
key 
Returns to the weight display without changing the number. 
 
2. When setting the number to 0, it is entered as 1 automatically. 
 
15.3.
15.3.
15.3.
15.3.    
Interval Data Output / Print
Interval Data Output / Print
Interval Data Output / Print
Interval Data Output / Print    
The AD-4405 can output and/or print data with a constant time interval using the 
internal calendar/clock function. The time intervals for serial data output and for the 
printer can be set individually from 1 second up to 23 hours and 59 minutes. 
Caution 
The interval time must be set to allow enough time to finish output or 
printing out. 
 
15. Common Printer and Data Output