A&D AD-4405 User Manual

Page of 80
 Page 
58 
AD-4405 Weighing Indicator
 
16.3.
16.3.
16.3.
16.3.    Command 
Format
Command Format
Command Format
Command Format    
Explanation of Command
Explanation of Command
Explanation of Command
Explanation of Commands
s
s
s    
 
  When performing a command, the received command or replay data is sent 
back. 
 
  When the received command can not be performed such as the "busy" state, the 
code " 
I " is sent back. In this case, insert a delay time. Electrical noise may be 
the cause of this error. 
 
  When receiving an undefined command (unknown command), a "?" is sent back. 
 
  When there is insufficient memory to store the data of the UFC commands, an 
“M” is sent back. 
 
  Optional addresses can be appended to a command. The address form is 
"@address" and the address is specified at F-Function F06. The reply (data or 
error code) is also sent with the address. 
Example: The command is "Display net data". The address is 23. 
 
  Sending a command 
Sending a command with address 
 
 
 
 
 
  Unperformed command reply 
Unknown command reply 
 
 
 
 
 
  Memory overflow (for UFC commands) 
 
 
 
 
 
16.3. Command Format 
Terminators 
Memory overflow error
Carriage return(0Dh) 
Line feed(0Ah) 
Unperformed command
Terminators 
Terminators 
Command 
Terminators 
Command 
N
M
@ 2 3
Address number 
Unknown command
Terminators 
?
@ 2 3
Address number