5-6. Counting mode (pcs) 
Weighs the sample pieces and calculate its unit weight. Using the sample unit weight, 
the scale counts the number of parts or items. 
‰
 As for the minimum unit weight acceptable, see the function setting “Umin”. 
Selecting the counting mode 
1. Press the    UNITS    key to select 
  . 
  
 
 
 
 
   
(  :pieces) 
Storing the sample unit 
2. Press the  SAMPLE  key to enter the sample 
unit weight storing mode. 
 
 
3.  To select the number of samples, press the  
1
SAMPLE    key. It may be set to 5, 10, 25, 50, 
or 100. 
 
 
 
 
4. Place a tare container on the weighing pan, and 
press the  RE-ZERO  key. Confirm that the 
right side of the number of samples shows zero. 
 
 
 
 
 
 
 
 
 
5. Place the number of samples specified on the 
pan. In this example, 25 pieces. 
 
6. Press the  PRINT  key to calculate and store 
the unit weight. Remove the sample. The 
balance is set to count objects with this unit 
weight. 
 
Counting the objects 
7. Place the objects to be counted on the pan. 
 
Counting mode using the ACAI function 
 ACAI
TM
 (Automatic Counting Accuracy Improvement) is a function that improves the 
accuracy of the unit weight by increasing the number of samples as the counting 
process proceeds. 
 
8. If a few more samples are added, the ACAI 
indicator is displayed. (To prevent an error, add 
three or more. The ACAI indicator will not be 
displayed if overloaded.) 
     0 
   10 - 
   25 - 
Container
Weighing pan
     25  
ACAI indicator
ACAI 
   
  
25  
Confirm the display
   25 0 
ZERO   
Each pressing switches 
the number of samples.
Samples
(25 pcs) 
To store