A&D EJ-1500 User Manual

Page of 42
 24
Data format   
 
S T 
0
0
0
0
.
0
0
  g C
R
  L
F
 
 
       Header  
 
  Data  
 
    Unit  Terminator 
 
 
 
‰
  There are four types of headers: 
 
ST : Stable weighing data (including % data) 
 
QT : Stable counting data 
 
US : Unstable weighing data (including count and %) 
 
OL : Out of weighing range (Over) 
 
‰
  The data is normally 9 digits including a decimal point and a sign. 
 
‰
  There are 14 types of units:   
 
  g  : Weighing data “gram”   
 
 P C  : Counting data “pcs” 
 
  %  : Percentage data “%” 
 
  o z  : Weighing data “decimal ounce” 
 
 
l b  : Weighing data “decimal pound” 
 
o z t  : Weighing data “troy ounce” 
 
 
c t  : Weighing data “carat” 
 
m o m  : Weighing data “momme” 
 
d w t  : Weighing data “penny weight” 
 
G N  : Weighing data “grain” 
 
  N  : Force data “Newton” 
 
 
t l  : Weighing data “tael” 
 
  t  : Weighing data “tola” 
 
DS 
: Calculated density (specific gravity) value 
 
‰
  The terminator is always C
R
L
F
 
‰
  Example of output data: 
 
Weighing  data  “gram” 
S T ,
+ 0 0 1 2 3 4
 
  g C
R
L
F
 
Counting data 
Q T ,
+ 0 0 0 1 2 3 4  5 
  P C C
R
L
F
 
Percentage data 
S T ,
+ 0 0 0 1 2 3
.  4 
 
  % C
R
L
F
 
Out of range “gram” (+) 
O L
,
+ 9 9 9 9 9 9
.  9 
 
  g C
R
L
F
 
Out of range “pcs” (-) 
O L
,
- 9 9 9 9 9 9 9  9 
  P C C
R
L
F
 
Data output mode 
‰
 Stream mode   
Set the function “prt 0”. 
The balance outputs the current display data. The data-update rate is approximately 
10 times per second. This rate is the same as the display-update. 
The balance does not output data while it is in the setting mode. 
Separator
(“ ” shows a space.)