SECTION 8 RS-232 OUTPUT                  
 
The QHW Series of scales can be ordered with an optional RS-232 output.     
 
Specifications: 
RS-232 output of weighing data 
ASCII code 
4800 Baud 
8 data bits 
No Parity 
 
Connector:    9 pin d-subminiature socket 
 Pin 
 
Output 
 
Pin 3    Input,    not used at this time 
 
Pin 5  Signal Ground 
 
Data Format for normal weighing operations, parts counting or recalling of totals from 
memory will all be different.    Examples follow: 
 
Normal Output 
 
When percent weighing the output is the weight shown in percent only. 
GS    1.234kg            GS for Gross weight, NT for net weight and a unit of
weight 
No..             1 
      This number increments every time a new value is 
                                                      stored in memory 
Total  1.234kg              The total value stored in memory 
<lf>                        Includes 2 line feeds 
<lf> 
GS    100.00% 
    GS for Gross weight, NT for net weight and a unit of weight
<lf>                              Includes 2 line feeds 
<lf>