SECTION 5—WHEELS
39
NOTE: If replacing rear wheels with a different size than what was originally on the wheelchair, 
the front caster height must also be changed to keep the wheelchair frame parallel to the floor. Refer 
to Adjusting Front Caster Height on page 55 
ONLY after contacting Invacare. Anti‐tipper height 
(if applicable) must also be adjusted to maintain 1‐1/2 to 2‐inch clearance between bottom of the 
anti‐tipper wheels and the floor. Refer to Adjusting/Replacing Anti‐tipper on page 66.
Opening/Closing Camber Clamps
 WARNING
QUICK RELEASE LEVERS - Make sure the quick release levers are in the closed 
position before using the wheelchair, otherwise personal injury or damage to the 
wheelchair may occur.
STANDARD AND SUSPENSION CAMBER CLAMPS - Make sure the hex screws 
are securely tightened before using the wheelchair, otherwise personal injury or 
damage to the wheelchair may occur.
CAUTION
A4 CAMBER SYSTEM ONLY - DO NOT close the quick-release levers or tighten 
the socket screws or hex screws without camber inserts in the axle tube. Damage to 
the axle tube will occur.
NOTE: For this procedure, refer to FIGURE 5.6.
1. Perform one of the following to open a camber clamp:
• Quick Release Levers ‐ Pull the quick release lever to the open position.
• Standard Camber Clamps ‐ Loosen, but DO NOT remove the socket screw on the 
camber clamp.
• Suspension Camber Clamps ‐ Loosen, but DO NOT remove the hex screw on the 
bottom rear of the camber clamp.
2. Perform one of the following to close a camber clamp:
• Quick Release Levers ‐ Secure the quick release lever to the camber clamp by 
threading it completely into the camber clamp. Push the quick release lever on the 
camber clamps to the closed position.
• Standard Camber Clamps ‐ Securely tighten the socket screw to secure the axle 
tube.
• Suspension Camber Clamp ‐ Securely tighten the hex screw on the bottom rear of 
the camber clamp to secure the axle tube.