Siemens LR 460 User Manual
Page 44
SITRANS LR 460 – INSTRUCTION MANUAL
7ML19985JM01
mmmmm
Pa
ra
m
e
te
rs
Quick Start Wizard
The Quick Start wizard groups together all the settings you need to make for a simple
application.
application.
•
The wizard is a complete package and settings are inter-related.
•
Do not use the wizard to modify individual parameters.
•
You can access the Quick Start wizard either via SIMATIC PDM, or via the handheld
programmer.
programmer.
•
There is no need to perform a reset before a Quick Start.
1. Quick Start
In SIMATIC PDM, Open the menu Device – Quick Start, and follow steps 1 to 5.
If you are using the handheld programmer, activate program mode and work
through the Quick Start sequence.
through the Quick Start sequence.
1.1. Language
1.2. Application Type
1.3. Operation
Operation types
Options
ENGLISH, DEUTSCH, FRANCAIS, ESPANOL
Options
STEEL
Silo construction
CONCRETE
Options
LEVEL
Distance to material surface referenced from Low
Calibration Point (process empty level).
Calibration Point (process empty level).
SPACE
Distance to High Calibration Point (process full level)
referenced from material surface.
referenced from material surface.
DISTANCE
Distance to material surface referenced from Sensor
Reference Point.
Reference Point.
SITRAN
S LR 40
0
high
calibration
point
low calibration point
level
space
distance
sensor reference point
(flange face)