P6249 4 GHz 5X Active Probe Instruction Manual
39
Performance Verification
Use the following procedures to verify the warranted specifications
of the P6249 probe. Before beginning these procedures, refer to
page 47 and photocopy the test record and use it to record the
performance test results. The recommended calibration interval is
one year.
These procedures test the following specifications:
H Output offset voltage
H DC attenuation accuracy
H Rise time
Equipment Required
Refer to Table 7 for a list of the equipment required to verify the
performance of your probe.
Table 7: Equipment required for performance verification 
Item Description
Performance
Requirement
Recommended Example
Oscilloscope
TEKPROBE interface
Tektronix TDS 694C
Sampling Oscilloscope
Tektronix 11801
Sampling Head
Tektronix SD24
TEKPROBE Power Supply
Tektronix 1103
Calibration Step Generator
067-1338-0X
Adapter
Probe tip
015-0678-00
DC Power Supply
1 VDC at 1 mA
Tektronix PS280
DMM
0.5% accuracy at 2 VDC
Tektronix TX3
Feedthrough Termination
50
W ±0.05 W
011-0129-00