Fluke 1552A EX User Manual

Page of 22
 
Stik Thermometer 
 
Safety Information 
  Check for proper operation of the Product before taking it 
into a hazardous area. Should the Product not operate as 
expected do not take the Product into a hazardous area and 
return it to the manufacturer for inspection. 
  The Product is intended for installation only in locations 
providing adequate protection against the entry of solid 
foreign objects or water capable of impairing safety. 
W
Caution 
To prevent possible damage to the Product or to equipment 
under test: 
  If the display message changes to “OL,” the range limit is 
exceeded and the Product must immediately be removed 
from the heat source to prevent damage. 
  Do not immerse the Probe beyond its maximum immersion 
depth. This may damage its internal electronics. See 
“Immersion Depth and Usage” later in this manual. 
  DO NOT drop or strike the Probe in any way. Mechanical 
shock damages the Probe internally and affects it 
calibration. 
If the Product overheats or is exposed to sudden physical shock, examine it for damage that 
can cause a safety risk. If possible, compare the displayed temperature to a known reference 
before Product use. If not sure, send the Product to Fluke Corporation. Refer to “How to 
Contact Fluke”. 
Substitution of components will impair suitability for hazardous locations. 
ATEX Safety Information 
This manual contains data and safety regulations that must be followed for safe, reliable 
use of the Product in hazardous areas under the detailed conditions. If you do not follow 
these instructions, personal injury or Product damage can occur. Violation of applicable 
legislation can also occur. Read the full manual before you use the Product. To make sure 
of safe Product operation, fully follow all instructions and warnings in the manual. If you 
are not sure (due to translation and/or printing errors), refer to the English manual. 
An “Ex-hazardous area,” as used in this manual, refers to an area made hazardous by the 
possible presence of flammable or explosive vapors. These areas are also referred to as 
“hazardous locations.” 
Ex ib IIB T4 Gb (–10 °C 
ITS10ATEX27114X
Ex ib IIB T4 Gb 
II 2 G IECEx ITS10.0049                              0344
Manufactured by Martel Electronics, Inc., 
3 Corporate Park Dr.
Derry, NH, USA
≤ Ta ≤ +50 °C)
 
gjo006.eps