Fluke 1522 User Manual

Page of 60
3
Specifications and Environmental
Conditions
3.1
Specifications
Resistance Range
0
Ω to 500 kΩ
Resistance Accuracy, PRT, one
year
0
Ω to 25Ω: 0.002Ω
25
Ω to 400Ω: 0.008% (80 ppm) of reading
Resistance Accuracy,
thermistor, one year
0
Ωto 2 kΩ: 0.4Ω
2 k
Ω to 200 kΩ: 0.02% (200 ppm) of reading
200 k
Ω to 500 kΩ: 0.03% (300 ppm) of reading
Temperature Range
PRT: –200°C to 962°C (–328°F to 1764°F)
Thermistor: –50°C to 150°C (–58°F to 302°F)
Temperature Accuracy, PRT
–200°C to 100°C: ±0.025°C (±0.045°F)
100°C to 400°C: 0±.05°C (±0.09°F)
400°C to 800°C: ±0.1°C (±0.18°F)
800°C to 962°C: ±0.15°C (±0.27°F)
Temperature Accuracy,
2.25 k
Ω thermistor
–50 to 25°C: ±0.005°C (±0.009°F)
25°C to 50°C: ±0.01°C (±0.018°F)
50°C to 75°C: ±0.03°C (±0.054°F)
75°C to 100°C: ±0.08°C (±0.144°F)
Temperature Accuracy,
10 k
Ω thermistor
0 to 50°C: ±0.005°C (±0.009°F)
50°C to 75°C: ±0.01°C (±0.018°F)
75°C to 100°C: ±0.02°C (±0.036°F)
100°C to 125°C: ±0.05°C (±0.09°F)
125°C to 150°C: ±0.1°C (±0.18°F)
Temperature Accuracy,
100 k
Ω thermistor
0 to 50°C: 0.006°C (0.011°F)
50°C to 150°C: 0.009°C (0.016°F)
Resistance Resolution
0.001
Ω
Temperature Resolution
0.001°C, F, K, R
Probe
IEC-751 or DIN-43760 PRT
Callendar-Van Dusen calibrated PRT; nominal 100
Ω
ITS-90 calibrated 25
Ω or 100Ω PRT
YSI-400 series or equivalent 2252
Ω thermistor
Steinhart-Hart thermistor polynomial; nominal R(25°C) 2k
Ω to
100k
Ω
Probe Connector
Hart INFO-CON connector
Maximum Acceptable Probe
Lead Resistance
10
Ω
Probe Excitation Current
PRT: 0.5 mA
Thermistor: 5
μA
Measurement Period
1 second
9
3 Specifications and Environmental Conditions
Specifications