Fluke 1529 Chub-E4 User Manual

Page of 154
3
Specifications and Environmental
Conditions
3.1
Specifications
PRT
Thermistor
Thermocouple
Inputs
2 channels PRT/thermistor and 2 channels thermocouple
or 4 channels PRT/thermistor
or 4 channels thermocouple
PRT/thermistor channels accept 2,3, or 4 wires
Thermocouple channels accept B, E, J, K, N, R, S, T, and Au-Pt thermocouple types
Temperature Range
1
–189 to 960°C
–50 to 150°C
–270 to 1800°C
Measurement Range
0 to 400
Ω
0 to 500k
Ω
–10 to 100 mV
Resistance / Voltage
Accuracy
2
0 to 20
Ω : ±0.0005Ω
20 to 400
Ω: ±25 ppm of reading
0 to 5k
Ω: ±0.0005kΩ
5 to 200k
Ω: ±100 ppm of
reading
200 to 500 K
Ω: ±300 ppm of
reading
–10 to 50 mV: ±0.005 mV
50 to 100 mV: ±100 ppm of
reading
(Internal RJC: ±0.25°C)
Characterizations
ITS-90,
PT-100 (ASTM E1137, DIN 43760,
IEC-751),
Callendar-Van Dusen
Steinhart-Hart, YSI-400
NIST Monograph 175, 3-point
deviation function applied to
NIST 175, 6th-order polynomial
Temperature Accuracy (meter
only)
1
±0.004°C at –100°C
±0.006°C at 0°C
±0.009°C at 100°C
±0.012°C at 200°C
±0.018°C at 400°C
±0.024°C at 600°C
±0.0025°C at 0°C
±0.0025°C at 25°C
±0.004°C at 50°C
±0.010°C at 75°C
±0.025°C at 100°C
Type Ext./Int.RJC
B
±0.6°C/±0.6°C
E
±0.07°C/±0.25°C
J
±0.1°C/±0.35°C
K
±0.15°C/±0.4°C
N
±0.15°C/±0.3°C
R
±0.4°C/±0.5°C
S
±0.5°C/±0.6°C
T
±0.1°C/±0.3°C
Temperature Resolution,
typical
3
0.001°
0.0001°
0.01 to 0.001°
Operating Range
1
16 to 30°C
Measurement Interval
0.1 second to 1 hour; inputs may be read sequentially or simultaneously at 1 second or greater interval
Excitation Current
1mA
(reversing)
2 and 10
μA, automatically
selected
N/A
Display
1.3” x 5" backlit LCD graphical display
Display Units
°C, °F, K,
Ω, KΩ, mV
Data Logging
Up to 8,000 time- and date-stamped measurements can be logged
Logging Intervals
0.1, 0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes
Averaging
Moving average of most recent 2 to 10 readings, user selectable
Statistical Data
Average, standard deviation, minimum, maximum, spread, N, difference (delta)
9
3 Specifications and Environmental Conditions
Specifications