Fluke 1529 Chub-E4 User Manual
![Fluke](https://files.manualsbrain.com/attachments/9d57315a76f85bd9af23da46bf7ecaf7f248b87d/common/fit/150/50/7671c6748e9e70e4c5afb082a772388950496a1161e6cfdedf4b942e284b/brand_logo.png)
0
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).
8.4.2.6
SENSe<chn>:RESistance:RANGe <num>|MIN|MAX|DEF
Sets the resistance range setting.
SENS4:RES:RANG MIN
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). The <num> parameter must be a number between 0 and 1. Specifying
the MIN, MAX or DEF parameter sets the resistance range setting to the mini-
mum, maximum, or default value respectively.
to 14). The <num> parameter must be a number between 0 and 1. Specifying
the MIN, MAX or DEF parameter sets the resistance range setting to the mini-
mum, maximum, or default value respectively.
This command may be password-protected. To access this command:
• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).
8.4.2.7
SENSe<chn>:RESistance:WIRE? [MIN|MAX|DEF]
Returns the probe wire setting.
SENS1:RES:WIRE?
3
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.
1529 Chub-E4 Thermometer Readout
User’s Guide
User’s Guide
92