Operating The M703E over the Serial I/O Ports 
M703E Calibrator Operator’s Manual 
7.5.2.1. Manually Configuring the Network IP Addresses  
There are several circumstances when you may need to configure the interface settings of the calibrator’s 
Ethernet card manually.  The 
INET sub-menu may also be used to edit the Ethernet card’s configuration 
properties   
  Your LAN is not running a DHCP software package, 
  The DHCP software is unable to initialize the calibrator’s interface; 
  You wish to program the interface with a specific set of IP addresses that may not be the ones 
automatically chosen by DHCP. 
Editing the Ethernet Interface properties is a two-step process.   
STEP 1: Turn DHCP OFF:   While DHCP is turned ON, the ability to set INSTRUMENT IPGATEWAY IP and 
SUBNET MASK manually is disabled 
SETUP X.X            COMMUNICATIONS MENU
ID ADDR
INET
EXIT
SETUP X.X            ENTER PASSWORD:818
8
1
8
ENTR EXIT
SETUP X.X            DHCP:ON
<SET SET>
EDIT
EXIT
SETUP X.X            DHCP:ON
ON
ENTR EXIT
SETUP X.X            DHCP:OFF
OFF
ENTR EXIT
Toggle this key 
to turn
 DHCP 
ON/OFF
ENTR accepts 
the new setting
EXIT ignores the 
new setting
Continue to Step 2 Below
STANDBY            ACT =STANDBY