Teledyne 703E User Manual
TELEDYNE API
M703E Calibrator Operator’s Manual
Operating the M703E Calibrator
6.5.1.3. Setting the Repeat Count for a Sequence
The sequence can be set to repeat a certain number of times, from 1 to 100. It can also be set to repeat
indefinitely by inputting a zero (0) into the
indefinitely by inputting a zero (0) into the
REPEAT COUNTER.
To set th
e REPEAT COUNTER, press:
SETUP X.X REPEAT COUNT:1
<SET SET>
EDIT
EXIT
SETUP X.X NAME:[0]
0
0
1
ENTER EXIT
EXIT discards the
new NAME
ENTR accepts the
new NAME
Toggle these keys to set the repeat count from 1 to 100.
Enter “0” to cause the sequence to loop indefinitely
Continue pressing SET> until ...
STANDBY ACT =STANDBY
<TST
TST>
GEN STBY SEQ
SETUP
Make sure that the M703E
is in standby mode.
SETUP X.X
PRIMARY SETUP MENU
O3
SEQ
CFG
CLK PASS
MORE
EXIT
SETUP X.X
SEQUENCE CONFIGURATION
EDIT
PRINT
EXIT
SETUP X.X
END OF SEQUENCES
INS
PRNT
EXIT
SETUP X.X
1) SEQ [NAME], [X] STEPS
PREV NEXT
INS
DEL EDIT
PRNT
EXIT
SETUP X.X NAME:0
SET>
EDIT
EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown
in the message field
Scrolls back and forth between
existing sequences
This display only appears if there are no sequences currently
programmed into the M703E.
OTHERWISE ...
05744 Rev B
43