Appendix C: 
Scanner Specifications
Site Specifications
AC Power Requirements
1,750 watts peak
AC Frequency
50 or 60 Hz
AC Phase
Single
Operating Environment
Room temperature:
65 to 80° F (18 to 27° C)
Relative humidity:
20 to 70%
Ambient light:
less than 1000 lux; no exposure to direct lighting
Altitude:
7000 feet (2134 m) maximum
Scanner Specifications
Dimensions and Weight
Height:
15 in. (38,1 cm)
Width:
24 in. (61 cm)
Depth:
30 in. (76,2 cm)
Weight:
120 lbs (54,5 kg) 
Footprint: 5 square feet
Operating and Service 
Clearances
Rear:
1.5 feet (45,7 cm) for operation, 
3 feet (91,4 cm) for service
Sides:
2 feet (60 cm) for operation and service
Front:
4 feet (122,9 cm) for operation and service
Scanner Location
The scanner location relative to the host workstation is limited by the SCSI 
interface cable length; if you are using a cable stocked by Kodak, this 
distance is 2 meters or less. Longer cables are not supported for use with 
the scanner. The scanner must be the only SCSI device attached to the 
host computer. 
The scanner should be located on a table that:
is stable
can support approximately 120 lbs (54,5 kg)
is at least 30 in. (76,2 cm) long x 30 in. (76,2 cm) wide
Power Consumption 
and Thermal Load
Power Consumption: 425 watts maximum; 400 watts typical
Thermal Load:
1325 BTU operating; 1230 BTU standby
AC Voltage and Current 
Draw
100 V AC/50 Hz
4.2 amps operating;
 3.9 amps standby
100 V AC/60 Hz
4.1 amps operating;
3.85 amps standby
120 V AC/50 Hz
3.5 amps operating;
 3.2 amps standby
200 V AC/60 Hz
2.0 amps operating;
1.9 amps standby
200 V AC/50 Hz
2.0 amps operating;
1.85 amps standby
230 V AC/60 Hz
1.8 amps operating;
1.7 amps standby
250 V AC/50 Hz
1.75 amps operating;  1.65 amps standby