Kodak 2400DSV User Manual

Page of 131
7
Troubleshooting
116
7.6.11 S201: Interlock Switch
7.6.12 TS1: Thermostat
1
Is the circuit across COM and NC terminals 
conducting when the Switch is removed from 
the machine?
→Yes
Replace S201.
↓ No
2
Is the circuit across COM and NC terminals 
conducting when the Switch Actuator is 
depressed?
→No
Replace S201.
→Yes
OK
1
Is the Reset Pin in the depressed position?
→No
Depress the Reset Pin.
↓ Yes
2
Is the circuit across both terminals of the 
Thermostat conducting when the Thermostat 
is removed from the machine?
→No
Replace TS1.
→Yes
OK