Compact, lightweight, easy-to-use SLR luminance 
meters with a wide measuring range
LS-100
Main Features
Flareless SLR optical system for accurate 
measurements
The SLR (single-lens-reflex) optical system allows precise aiming 
and ensures that the viewfinder shows the exact area to be 
measured. The optical system is also virtually flareless, eliminating 
the influence of light from outside the measurement area.
Narrow acceptance angle for measurements 
of small specimens
Acceptance angles of only 1° for LS-100 and 1/3° for LS-
110 allow accurate measurements of small specimen areas. 
In addition, optional close-up lenses can be used to measure 
areas as small as ø1.3 mm when using LS-100 and ø0.4 
mm when using LS-110.
User calibration and color-correction 
functions
To increase the versatility of the LS-100 and LS-110
both models are equipped with user calibration and color 
correction functions. The user calibration function allows the 
meter to be calibrated to a user-selected standard instead 
of the preset Konica Minolta standard; this function can also 
be used to standardize the response of several meters. The 
color correction function allows the response of the meter to 
be adjusted when measuring colored specimens.
Luminance ratio and peak luminance 
measurements
In addition to measurements of the present luminance, the LS-
100 and LS-110 can also determine the percent ratio of the 
measured luminance to a luminance value stored in memory 
as well as the peak luminance or luminance ratio measured.
RS-232C data communication   
Use of the built-in RS-232C interface allows the meter to be 
connected to a personal computer.
Lightweight, compact design powered by a 
single 9V battery for portability
Luminance Meter LS-100
1° acceptance angle, 
Measuring range: 0.001 to 299,900 cd/m
2
(0.001 to 87,530 fL)
Luminance Meter LS-110
1/3° acceptance angle, 
Measuring range: 0.01 to 999,900 cd/m
2
(0.01 to 291,800 fL)
Relative Spectral Response
Reduction of Flare
Ideally, the relative spectral responsivity of the luminance meter 
should match V (λ) of the human eye for photopic vision.
As shown in the graph above, the relative spectral 
responsivity of Konica Minolta Luminance Meters LS-100/
LS-110 is close to the CIE spectral luminous efficiency V (λ).
CIE ; Commission Internationale de I«Eclairage
f
1
’(CIE«s symbol) ;  The degree to which the relative spectral responsivity 
matches V (λ) is characterized by means of the error f
1
’. 
The degree to which the influence of light from outside 
the defined measuring area is eliminated is an important 
factor in the performance of luminance meters. In Konica 
Minolta Luminance Meters, the flare factor is kept to below 
1.5%, even if an object with 
extremely high luminance 
is just outside the meter’s 
measuring area. 
The graph at right shows the 
effect when a bright point 
is moved from A inside the     
measuring area to B just 
outside the measuring area.
If the measured value at 
A is defined at 100%, the 
measured value at B would be 
less than 0.1%.
100%
10%
1%
0.1%
30' 15' 0 15' 30'
A
B
Angle of
acceptance
in minutes
1°(1/3°) measuring spot
Relative sensitivity
(10 )
' (5 )
'
(5 )
' (10 )
'
750
700
650
600
550
500
450
400
0
10
20
30
40
50
60
70
80
90
100
Relative sensitivity (%)
Wavelength (nm)
Konica Minolta Luminance
Meters LS100/LS-110
The spectral
luminous efficiency
 (λ)
Angle Finder V
N
 allows the 
measuring area and 
measurement display inside the 
viewfinder to be seen at an angle 
of 90° to the normal viewfinder 
optical axis. Angle Finder V
N
 can 
also be focused and the 
magnification can be set to 1x or 
2x.
 ⿎
Specifications 
Model
Luminance Meter LS-100
Luminance Meter LS-110
Type
SLR spot luminance meter for measuring light-source and surface brightness 
Measuring angle
1/3°
Optical system 
85 mm f/2.8 lens; SLR viewing system; flare factor less than 1.5%
Angle of view
Focusing distance
1014 mm (40 in.) to infinity
Minimum measuring area
ø14.4 mm 
ø4.8 mm
Receptor
Silicon photocell
Response time
FAST:  Sampling time: 0.1s, time to display: 0.8 to 1.0s; SLOW:  Sampling time: 0.4s, time to display: 1.4 to 1.6s
Luminance units
cd/m
2
 or fL (switchable)
Measuring range
FAST  : 0.001 to 299,900 cd/m
2
 (0.001 to 87,530fL)
SLOW  : 0.001 to 49,990 cd/m
2
 (0.001 to 14,590fL)
FAST  : 0.01 to 999,900 cd/m
2
 (0.01 to 291,800 fL)
SLOW  : 0.01 to 499,900 cd/m
2
 (0.01 to 145,900 fL)
Accuracy*
1
0.001 to 0.999 cd/m
2
 (or fL): ±2% ±2 digits of displayed value
1.000 cd/m
2
 (or fL) or greater: ±2% ±1 digit of displayed value
0.01 to 9.99 cd/m
2
 (or fL): ±2% ±2 digits of displayed value
10.00 cd/m
2
 (or fL) or greater: ±2% ±1 digit of displayed value
Repeatability*
2
0.001 to 0.999 cd/m
2
 (or fL): ±0.2% ±2 digits of displayed value
1.000 cd/m
2
 (or fL) or greater: ±0.2% ±1 digit of displayed value      
0.01 to 9.99 cd/m
2
 (or fL): ±0.2% ±2 digits of displayed value
10.00 cd/m
2
 (or fL) or greater: ±0.2% ±1 digit of displayed value
Temperature/humidity drift
Within ±3% ±1 digit (of value displayed at 20°C/68°F) within operating temperature/humidity range
Calibration mode
Minolta standard/user-selected standard (switchable)
Color correction factor
Set by numerical input; range: 0.001 to 9.999
Reference luminance
1; set by measurement or numerical input
Measurement modes
Luminance; luminance ratio; peak luminance or luminance ratio
Display
External:      4-digit LCD with additional indications
Viewfinder:  4-digit LCD with LED backlight
Data communication
RS-232C; baud rate: 4800 bps
External control
Measurement process can be started by external device connected to data output terminal
Power source
One 9 V battery; power can also be supplied by optional Data Printer DP-10
Power consumption
While measuring button is pressed and viewfinder display is lit: 16 mA average
While power is on and viewfinder display is not lit: 6 mA average
Operating temperature/humidity range 0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Storage temperature /humidity range -20 to 55°C, relative humidity 85% or less (at 35°C) with no condensation
Dimensions
79x208x150 mm (3-1/8x8-3/16x5-7/8 in.) 
Weight
850 g (30 oz.) without battery
Standard accessories
Lens cap; Eyepiece cap; ND eyepiece filter; 9 V battery; Case 
*
1 Standard Illuminant A measured at ambient temperature of 20 to 30ºC
*
2 Standard Illuminant A
Optional Accessories
 ⿎
System diagram
 (Optional Accessories)
Close-Up Lenses
No. 153
No. 135
No. 122
No. 110
Angle Finder  V
N
PC
Connecting Cable
LS-A11
(for data output)
Connecting Cable
LS-A12
(for data communication)
Minimum measuring area
Close-Up
Lenses
With
LS-100
With
LS-110
No.153 ø8.0  mm ø2.7  mm
No.135 ø5.2  mm ø1.8  mm
No.122 ø3.2  mm ø1.1  mm
No.110
ø1.3  mm ø0.4  mm
(Theoretical values)
Close-Up Lenses
Angle Finder V
N
Luminance Meters
 
 LS-100/LS-110
Luminance Meters LS-100/LS-110
Luminance Meters LS-100/LS-110